On Line Gas Analysers
 
 
 
 
 
 

Oxygen Analyser Digital, Model OA-12, 0 - 100% (92x96x300mm)

Artech offers high performance low cost Analysers for measuring oxygen levels from pure oxygen down to fractional percentage levels. These analysers are based on electrochemical cell sensors. Model OA-12 can be used from 0-100% oxygen. Digital readout of concentration and dual alarms are standard features. Each alarm channel is provided with a set of changeover relay contacts. The alarm set points can be set and displayed from the front panel. A needle valve regulator and a flow indicator enable the user to regulate sample gas flow from the front panel.
Applications include

• Air separation plants
• Cylinder gas quality
• Fruit storage
• Food packing machines

• Medical monitoring
• Purge gas monitoring
• Landfill gas
• Nitrogen purged soldering systems

Specifications
Range : 0 - 100 %
Resolution : 0.1 %
Indication : 4 Digit LED Display, 1/2" height
Accuracy : + 1 % at constant temperature and pressure
Response Time of Sensor : 90 % in 15 seconds
Transport time from
inlet port
:

Less than 3 seconds at recommended
flow-rate of 1 litre/minute

Flow Rate Indication on front panel : 0 - 1 litre/minute
Flow Rate Adjustment :

By needle valve control on the front Panel

Sample Gas Temperature : 5 - 40 o C at inlet port
Sample Gas Pressure : 1.2 Kg.(ab.) at inlet port
Calibration : With atmospheric air
Alarm :

High & Low Alarms adjustable throughout the range

Alarm Output :

One set each of relay changeover contacts rated at 5 Amps(resistive)

Power Supply : 230V AC + 10%
Dimensions : W x H x D : 192 x 96 x 288 mm
Panel Cutout: 186 x 92 mm