| Range |
: |
0-1000 ppm Oxygen or 0-30% automatically selected |
| Indication |
: |
4 Digit LED Display / 1/2" height |
| Accuracy |
: |
+ 1% of F.S. + 1 count at constant temperature and pressure |
| Type of Sensor |
: |
Electrochemical cell (Two cells - one for ppm, one for percentage) |
Response Time of
Sensor (T-90) |
: |
1 minute. (Sample transport time extra initially 6 hours from exposure to atmosphere (ppm only)
|
Transport time from
inlet port |
: |
Less than 3 seconds at recommended flow-rate of 1-litre/minute. |
| Flow Rate Indication
on front panel |
: |
0-1 litre/minute |
| Flow Rate Adjustment |
: |
By needle valve control on front panel |
| Sample Gas Pressure |
: |
1.2 kg/cm2 (ab.) at inlet port. |
| Calibration |
: |
With ppm calibration gas and ambient air |
| Alarm Output |
: |
One set of changeover contacts rated at 5 Amps(resistive) for each alarm |
| Dimensions |
: |
288 x 144 x 220mm (W x H x D) |
| Output signal |
: |
4-20 mA - 500 ohms maximum load
For ppm & percentage range |
| Specification, Configuration & Dimensions subject to change |